SKU PEAK-INTURA-H2-100
Intura H2 100 produces 99.99999 % hydrogen at up to 100 cc/min and 175 psi for GC and GC-MS carrier and detector gas, ICP-MS and hydrogenation, using the latest PEM cell technology with water-quality monitoring and pre- and post-filtration. Capacity can be expanded by daisy-chaining generators.
Prices are given by quotation. Add this to your quote cart to request one.
| Part number | 3305698-0-01 (auto-fill) / 3305698-0-11 (pressure-fill) |
|---|---|
| Max flow | 100 cc/min |
| Output pressure | 175 psi / 12.1 bar |
| Purity | 99.99999 %, phthalate and BHT free |
| Dew point | -75 °C |
| Power consumption | 296 W |
| Electrical | Universal 100-240 V 50/60 Hz |
| Water | < 0.1 µS/cm (ASTM Type II or approved DI purifier) |
| Dimensions (H × W × D) | 324 × 330 × 520 mm |
| Weight | 25 kg |
| Noise | < 55 dB(A) |
| Max altitude | 3000 m |
| Warranty | 12 months (extendable to 5 years); PEM cell 36 months |
| Maintenance | Annual DI cartridge change |